Source scpi commands – AMETEK ReFlex Programming Manual User Manual
Page 123
Load Module Remote Programming
ReFlex Power™ Programming Manual
M380056-03 Rev L
119
5.9
SOURCE SCPI COMMANDS
5.9.1
SOURCE
SCPI
C
OMMAND
S
UMMARY
SOURce
:CURRent
:CURRent?
:PROTection
:PROTection?
:DELAY
:DELAY?
:RESistance
:RESistance?
:VOLTage
:PROTection
:PROTection?
:POWer
:PROTection
:PROTection?
:DELAY
:DELAY?
5.9.2
SOURCE
SCPI
C
OMMAND
R
EFERENCE
Command
Description
SOURce
SOURCE subsystem of commands
:CURRent
Sets the input DC current setpoint in Amps. The range of the current setpoint
is 0A to 30.1A in high power units and 0A to 15.05A in low power units.
Current setpoint takes effect as soon as the command is sent. The Reset
command sets the current setpoint to the default value of 0A.
:CURRent?
Returns the most recently set current setpoint.
:PROTection
Set the overcurrent protection limit. The range of the current protection
setpoint is 5% to 105% of 30.0A for high power units, and of 15.0A for low
power units. A new over current protection limit takes effect as soon as the
command is sent. The Reset command resets the protection level to 105%.
:PROTection?
Returns the most recently set overcurrent protection limit
:DELAY
Sets the over current protection delay timer in millisecond units. The range for
the delay time is 1.00 to 100.00mSec, in 10uSec increments. Delay time
takes effect as soon as the command is executed. *RST resets the delay
time back to the default value of 100mSec.
:DELAY?
Returns the most recently set over current protection delay time.
:RESistance?
Returns the most recently set resistance setpoint.
:RESistance
Sets the input resistance setpoint, range from 1 to 5,000
Ω
’s.
:VOLTage
Input voltage protection.
:PROTection
Voltage protection subsystem commands
:PROTection?
Returns the most recently set over voltage protection limit
:POWer
Power Protection subsystem commands
:PROTection
Set the Power protection limit
:PROTection?
Query the Power protection limit