Common scpi commands – AMETEK ReFlex Programming Manual User Manual
Page 60
ReFlex Power™ Programming Manual
DC Module Remote Programming
56
M380056-03 Rev M
REPLY
DESCRIPTION
Unable to change Output State
OUTP
[to power on or off
the DC module
Relay hot-switching prohibited or Parallel
Slave
OUTP
[to power on or off
the DC module
Unable to close Default Sense Relay, Default
Isol Relay is open
User tries to close the sense relay with the
CAL
command when the isolation relay is open
Unable to open Output Relay, Sense Relay is
closed
User tries to open the isolation relay with the
CAL
command when sense relay is closed
Relay hot-switching prohibited
User tries to open the isolation relay with the
OUTP
command when the
output voltage is on
Unable to change Polarity, an Output or
Sense relay is closed
User tries to change the polarity of the module
with the
OUTP
command
when Isolation and sense relay is closed
Unable to close Sense relay, Output Relay is
NOT closed
User tries to close the sense relay with the
OUTP
command when the
isolation / output relay is not closed
Module Firmware too old to support this RFP
Controller Firmware Revision
The addressed module’s firmware version is too
old to be compatible with the RFPC’s firmware
3.5
COMMON SCPI COMMANDS
The following commands are common to all SCPI instruments and declared
mandatory by IEEE-488.2. In the following table, the DCPS is defined as the
“device” on the bus.
Table 3-2. Common SCPI Commands
Command
Description
*CLS
Clears module hardware faults latch
*IDN
Returns the device identification as an ASCII string.
Response:
Example: ELGAR,RFP-D1016-021,1234A56789,3.000.001
*OPC
Operation Complete Query: Returns the integer value “1” when all pending
operations are complete.
*RST
Resets the supply to its Power ON (PON) state.
Clears status reporting data structures the Protection Event and the Condition
Status Registers. Enable registers are not cleared by this command. RST will reset
all modules.
*TST
Initiate self-test
*TST
Returns the results from the last self test the module performed. Returns a decimal
value, convert to Hex because more than one bit can be set. See Table 2-6