AMETEK ReFlex Programming Manual User Manual
Page 131
Fixed Power DC Remote Programming
ReFlex Power™ Programming Manual
M380056-03 Rev L
127
6.5.2
CALIBRATION
SCPI
C
OMMAND
R
EFERENCE
Command
Description
CALibrate
1
CALIBRATE sub-commands
:UNLock
Enables the EEPROM memory for storing calibration constants.
The enable string is <“6867”>.
:STORe
Stores the calibration constants in the EEPROM memory.
:LOCK
Disables access to the EEPROM memory, thereby preventing
storing of calibration constants.
:POSitive15 <0+NR1>
Calibration for Positive 15 DC supply. Value is between 0 to 255
:POSitive15?
Query the pot setting
:POSitive5 <0+NR1>
Calibration for Positive 5 DC supply. Value is between 0 to 255
:POSitive5?
Query the pot setting
:NEGative15 <0+NR1>
Calibration for Negative 15 DC supply. Value is between 0 to 255
:NEGative15?
Query the pot setting
:NEGative5 <0+NR1>
Calibration for Negative 5 DC supply. Value is between 0 to 255
:NEGative5?
Query the pot setting
:MODule
:LASTCAL
Set the last calibration date
:LASTCAL?
Query the last calibration date
:NEXTCAL
Set the next calibration date
:NEXTCAL?
Query the next calibration date.
1
Calibration is required to be run twice, once with Sense relays closed and with Sense
relays open, then Store the calibration constants.