Measure scpi commands – AMETEK ReFlex Programming Manual User Manual
Page 121

Load Module Remote Programming
ReFlex Power™ Programming Manual
M380056-03 Rev L
117
Command
Description
:PROTection
Set the initial undervoltage protection value
:PROTection?
Return the initial undervoltage protection value
:VOLTage
:PROTect?
Get the initial over voltage protection value
:PROTect
Set the initial over voltage protection value
:INPut
Calibrate Input subsystem
:CURRent
Calibrate Input Current subsystem
:COUNTS
Enter the input current counts
:CALCulate
Calculate the current calibration
:FIVEPOINT <1-5>
Set the input current value for each calibration point
:
RANGE
Change the current range
:CURRent <1-2>
Set the input current calibration points
:CONDuctance
Calibrate Conductance subsystem
:COUNTS
Set the conductance counts
:CALCulate
Calculate the conductance calibration
:VOLTage <1-2>
Set the input voltage calibration points
:MEASure
Calibrate Measure subsystem
:OFFSET
Calibrate Measure Offset subsystem
:CALCulate
Calculate the voltage calibration
:UNLock
Unlock the calibration data file; enables storage of the calibration
constants. The enable string is <“6867”>.
:STORe
Stores the calibration constants.
:LOCK
Locks the calibration data file; disables calibration data store, thereby
preventing storage of the calibration constants.
5.7
MEASURE SCPI COMMANDS
5.7.1
MEASURE
SCPI
C
OMMANDS
S
UMMARY
MEASure
:CURRent?
:VOLTage?
:
POWer?
5.7.2
MEASURE
C
OMMANDS
R
EFERENCE
Command
DESCRIPTION
MEASure
MEASURE subsystem of commands
:CURRent?
Returns the floating-point value of the DC input current.
:VOLTage?
Returns the floating-point value of the DC input voltage.
:
POWer?
Returns floating point value of DC power