AMETEK ReFlex Programming Manual User Manual
Page 59

DC Module Remote Programming
ReFlex Power™ Programming Manual
M380056-03 Rev L
55
The
*CLS
command will clear all entries from the queue.
The following error codes are defined in the SCPI 1999.0 specification and
are supported by the DCPS. Error codes are in the range of [-32768, 32767].
SCPI reserves the negative error codes and 0, while error codes greater than
0 are device specific errors.
0, No error
The error queue is empty.
-102, Syntax error
An unrecognized command or data type was encountered.
-200, Execution error
This is a generic error when a specific error could not be identified. An
execution error could be the result of the following condition:
A command could not be executed due to the current operational
condition.
-222, Range Error
A
150V, when the maximum allowed value is 135V.
-350, Queue overflow
The Error/Event Queue can contain up to 10 entries. If more than 10
error/event conditions are logged before the SYSTem:ERRor? query, an
overflow will occur; the last queue entry will be overwritten with error -350.
When the queue overflows, the least recent error/event remains in the queue
and the most recent error/event is discarded.
NOTE: Some SCPI commands will generate the following replies, which can
be viewed with the SYST:ERR? command:
REPLY
DESCRIPTION
Invalid 5PCalData
Error occurs at the boot up sequence if the 5-point
calibration data are not valid.
Unable to open Module Data file - file system
problem
Error occurs at the boot up sequence if the
module data file does not open
Invalid Password.
CAL
“Password” [to unlock the
files to store the calibration or module data, but
the password used is incorrect]
Unable to open Module Data file - file is
locked
The
CAL
command is not used
before using
CAL
command
Unable to change Output State - Output
Inhibited
OUTP
[to power on or off
the DC module