ReFlex Power™ Programming Manual

DC Module Remote Programming

58

M380056-03 Rev M

:LOCK

:STORe

:UNLock

:REVERT:FACTory

3.6.2

CALIBRATION

SCPI

C

OMMAND

R

EFERENCE

Command

Description

CALibrate

Calibration subsystem. Module ‘n’.

:INITial

:CURRent

[:AMPLitude]

Sets the power-on default value of current.

[:AMPLitude]?

Query the power-on default value of current

:PROTection

Sets the power-on default value of the over current
protection.

:PROTection?

Query the power-on default value of the over current
protection.

:VOLTage

[:AMPLitude]

Sets the power-on default voltage. Should be set to zero
volts to avoid hot switching of output relay.

[:AMPLitude]?

Query the power-on default voltage. Should be set to zero
volts to avoid hot switching of output relay.

:PROTection

Sets the power-on default value of the over voltage
protection.

:PROTection?

Query the power-on default value of the over voltage
protection.

:STATe

1

Set the initial POR output state of the unit

:STATe?

Query the initial POR output state setting of the unit

:UNDERVOLTage

:PROTection

Sets the power-on default value of the under voltage
protection.

:PROTection?

Query the power-on default value of the under voltage
protection.

:DEFAult

Sets initial settings to Factory defaults of 0.0
Volts/Amps/UVP, OVP/OCP to Module Max values and
POR State to Off.

:OUTPut

:CURRent

:COUNTS

Set Current DAC setpoint, in counts

:FIVEPOINT<1-5> <0+NRf>

Sets the current output calibration point (1 through 5). The
actual output current is measured with an external meter.

:FIVEPOINT?

Returns 5 pairs of current calibration setpoint count and
related current values

:PROTection

:COUNTS

Set Protection value 0-4095

:VOLTage

:COUNTS

Set Voltage DAC setpoint, in counts

:FIVEPOINT<1-5>

Sets the voltage output calibration point (1 through 5). The
actual output voltage is measured with an external meter.