beautypg.com

Standing vacuum test – Carrier 19XL User Manual

Page 131

background image

b. If the machine fails this test, check for large leaks (Step 2b).

c. Dehydrate the machine if it passes the standing vacuum test. Follow the procedure in the

Machine Dehydration section. Charge machine with refrigerant (see Pumpout and

Refrigerant Transfer Procedures,

Machines with Storage Tanks

section, Steps 1a-e).

7. If a leak is found, pump the refrigerant back into the storage tank, or if isolation valves are

present, pump into the non-leaking vessel (see

Pumpout and Refrigerant Transfer

Procedures

section).

8. Transfer the refrigerant until machine pressure is at 18 in. Hg (40 kPa absolute).

9. Repair the leak and repeat the procedure, beginning from Step 2h to ensure a leaktight

repair. (If machine is opened to the atmosphere for an extended period, evacuate it before

repeating leak test.)

Standing Vacuum Test

When performing the standing vacuum test, or machine dehydration, use a manometer or a

wet bulb indicator. Dial gages cannot indicate the small amount of acceptable leakage during a

short period of time.

1. Attach an absolute pressure manometer or wet bulb indicator to the machine.

2. Evacuate the vessel (see

Pumpout and Refrigerant Transfer Procedures

section) to at least

18 in. Hg vac, ref 30-in. bar (41 kPa), using a vacuum pump or the pumpout unit.

3. Valve off the pump to hold the vacuum and record the manometer or indicator reading.