Installation 6-29

G1

G2

G3

G4

G5

outer shield

supply +

UE-GND

RxTx -

RxTx +

FlowShark Pulse

white

green

blue

red

black

DataPro

2 x 1 12 V / 12 V

1p

2p

1

2

DataPro

2 x 1 24/24 Tr

1p

2p

1

2

1.5 mm min.

2

1.5 mm min.

2

-0.3 Tr (N)

Figure 6-39 Overvoltage protection for the combi sensor supporting water-ultrasonic measurement

D1

D2

D3

D4

D5

outer shield

supply +

UE-GND

RxTx -

RxTx +

FlowShark Pulse

white

green

blue

red

black

DataPro

2 x 1 12 V / 12 V

1p

2p

1

2

DataPro

2 x 1 24/24 Tr

1p

2p

1

2

1.5 mm min.

2

1.5 mm min.

2

-0.3 Tr (N)

air ultrasonic sensor

Figure 6-40 Overvoltage protection for the air-ultrasonic sensor

Please observe the non-reversed connection of the DataPro (p-side
to transmitter) and a correct, straight wiring supply. Ground
(earth) must lead to the unprotected side.

Please note that incorrect wiring will deactivate the overvoltage
protection function.