beautypg.com

Paramet.niv, paramet.txt, Parameter menu: communication, Remote control – ADS Environmental Services FlowShark Pulse QR 775004 A1 User Manual

Page 144

background image

8-54 ADS FlowShark Pulse O&M Manual

If the location name saved to the memory card is not consistent with the name of the
measurement location saved to the FlowShark Pulse, the unit will prompt the user to
format the card.

The FlowShark Pulse will create a new file for the readings with the new name,
provided the card does not require formatting.

PARAMET.NIV, PARAMET.TXT

These files are created when parameters are saved to the memory card.
PARAMET.NIV is required for uploading data to FlowShark Pulse.
PARAMET.TXT is the printable (text) version of PARAMET.NIV.

Use memory cards purchased only through ADS. Other
manufacturer’s cards may lead to irreversible loss of data,
measurement failure, or permanent transmitter reset.

Do not format memory cards on a PC. Format cards only on the
FlowShark Pulse. The FlowShark Pulse can neither use nor accept
cards that have been formatted on a PC.

Parameter Menu: Communication

Modifications in this menu are required only when the user will be accessing the unit
over the Internet or a local network.

The communication options available (local network or GSM/GPRS gateway) depend
on the type of transmitter (see Chapter 4, System Overview and Operation).

Figure 8-85 Options for Internet connection

Remote Control

Select one of the following options for remote access:

z

Disabled – no remote access available

z

Modem – (For the FlowShark Pulse, this option currently is unavailable.)