8-20 ADS FlowShark Pulse O&M Manual
Sensor
No.
Air-Ultrasonic
01
Depth measurement using an air-ultrasonic sensor
connected directly to the FlowShark Pulse.
Requires an air-ultrasonic sensor.
Water-Ultrasonic
02
Standard version. Flow velocity and depth
measurement using combi sensor from the bottom
up.
Requires a 2K combi sensor.
External Sensor
03
Second standard version. Depth measurement using
an external, 2-wire sensor powered by FlowShark
Pulse or by external measurement unit and depth
input via mA input signal.
Requires an external 4-20 mA sensor.
Fixed Value
04
Use this point to set parameters for pipes and pipes
that remain full. These applications typically do not
require depth measurements. Enter the consistent
depth level under Scale/Height.
Pressure
05
Depth measurement determined by a sensor-
integrated pressure measurement cell.
Requires a 3K or 4K combi sensor.
Pressure + Air-
Ultrasonic
06
Depth measurements performed using a sensor-
integrated pressure measurement cell and an air-
ultrasonic sensor connected directly to a FlowShark
Pulse.
Requires a 3K or 4K combi sensor and an air-
ultrasonic sensor.
Pressure + Water-
Ultrasonic
07
Depth measurements performed using combination
of sensor-integrated pressure measurement cell and
water-ultrasonic from bottom up.
Requires a 3K or 4K combi sensor.
Pressure + External
Sensor
08
Depth measurements performed using combination
of sensor-integrated pressure measurement cell and
external 2-wire sensor powered by FlowShark Pulse
or by an external measurement unit and depth input
via mA input signal.
Requires a 3K or 4K combi sensor and an external
4-20 mA sensor.