11-2 ADS FlowShark Pulse O&M Manual

Error

Possible Reason

Correction

Check depth measurement parameter Fixed
value
when measuring in a full pipe without
obtaining a depth measurement.

Transmitter

Check error memory log and take corrective
measures. Check cables, plug, and socket
connections and sensor installation.

Contact ADS if an CPU Error displays.

Examine the moment of failure through the
RUN > Trend menu.

Negative Flow Direction Check orientation of sensor installation and

then reposition sensor, if necessary.

For occasional reverse flows that are
followed by measurement failure, set the
minimum value to -3.28 feet per second (–
1.0m/s) in the CAL-Flow vel. – min. + max.
value
menu.

Configuration

Check the transmitter parameter settings
thoroughly.

Connection

Check power supply connection.

Check voltage of power supply.

Check the switch position on the connection
board.

Power Supply

Compare the power supply (AC or DC)
against the transmitter type.

Confirm that an ADS-supplied memory card
is in use. Replace card from unauthorised
manufacturer with ADS card, when
necessary.

Confirm memory card does not exceed
capacity requirements. Use cards only
within the maximum memory capacity
allowed.

Blank or Flickering
Screen

Memory card

Verify method used for formatting cards.
Send cards formatted on a PC to ADS.