4.3 Systems

761 Compact IC

81

Ident

User defined identifier (title) for the chroma-

togram to be displayed in the title bar of the

chromatogram window and in the

Chromatogram

open

window.

Calibration level

Calibration level (0 = sample; 1

n = calibra-

tion solutions).

Info 1 / Info 2

Sample

description.

Volume

Injected volume in

µ

L.

Dilution

Dilution of the sample.

Vial number

Autosampler vial position to take sample from.

Amount

Sample amount. If this value is different for the

calibration run (c) and the sample run (s), the

component concentrations of the sample are

calculated as follows:

C

s

= C

c

Amount

s

/

Amount

c

Internal standard amount

Concentration of the internal standard compo-

nent for relative concentration calculations.

Date/time when...

Date and time of sample collection (the default

values are equal to the date and time when the

chromatogram starts).

The

Verify sample

option is disabled if determinations are made with an

active sample queue.