3 method setup – Metrohm 761 Compact IC User Manual
Page 114
4 Operation
761 Compact IC
104
4.4.3 Method
setup
761 COMPACT IC / Method / Method setup
This menu item opens the
Method setup
window which includes the
most common parameters for data acquisition and evaluation of the
method and consists of the following tabs:
General
General information on determination.
Measure
Important data acquisition parameters.
Filters
Parameters for noise filtration.
Processing
Set actions that are performed when the chroma-
togram finishes.
Math
Parameters that are used for various types of cal-
culations.
General
Tab
General
of
Method setup
window with general description of the
method and determination. This tab is identical to the
General
tab called
with
Method / Passport
(see section 4.4.2).
Measure
Tab
Measure
of
Method setup
window with data acquisition parameters.
Status
Chromatogram measurement status (read-
only).
Sampling rate
Sampling rate for recording the chromatogram
(read-only; only displayed correctly for running
or recorded chromatograms).
Frequency divisor
Divisor to reduce the sampling rate.
Entry range:
1
…
9999
Start delay
Time delay before starting data acquisition in
min.
Detector
Detector
name.
Data acquisition source
Measuring channel (read-only).