3 Operating tutorial

761 Compact IC

58

11 Fill sample loop

Immerse the aspirating tubing 4 attached to connection 3 in

the standard solution.

Using the syringe fixed to connection 2 siphon in ca. 1 mL

standard solution.

12 Switch injection valve to "INJECT" position

Click on the

button of the 761 icon in the system

window to switch the injection valve to the "INJECT" position.

At the same time, the data recording is started automatically.

The status bar of the chromatogram window shows the mes-

sage

Measure

, the

SYSTEM STATE

window shows the mes-

sages

Running [761 Compact IC]

and

INJECT done [Suppressed]

.

When the analysis time of 24 min set in the method has

expired the recorded chromatogram is automatically evalu-

ated and stored. The chromatogram window then closes.

13 Open chromatogram

Click on

or

File / Open / Chromatogram

in the main win-

dow. Select the chromatogram file

*.chw

just recorded and

click on

. The chromatogram window is opened where

the found peaks are numbered and the baselines are drawn

in.

14 Modify integration parameters

Click on

or select

Integration

from the

Method

menu in

the main window to open the integration parameters window.

If required (e.g. if the fluoride peak is not detected), change

the time delay

Delay

before starting peak integration.