4 Operation
761 Compact IC
158
Started
Indication whether the sample has been started
or not:
0
sample not started
1
sample started
2
…
n
number of injections done
Level
Calibration level (see section 4.4.5) for the
sample: Level
0
stays for normal analysis, levels
1
and greater for calibration runs. Correct filling
of this column enables to calculate calibration
coefficients automatically.
Volume
Injected volume of sample in
µ
L.
Amount
Sample amount. If this value is different for the
calibration run (c) and the sample run (s), the
component concentrations of the sample are
calculated as follows:
C
s
=
C
c
•
Amount
s
/
Amount
c
Shut down system after the queue finishes
If this option is enabled, the system hardware is
shut down after the sample queue has been fin-
ished (all pumps are switched off), but it re-
mains connected.
Start execution of the sample queue from the
first line with
Started = 0
.
Stop execution of the sample queue after the
current determination has been finished.
Open the sample queue editor program for ed-
iting the sample queue table (see section
4.6.3).
Close the sample queue overview table.
Start sample queue
SAMPLE QUEUE OVERVIEW /
Control / Start
Start execution of the sample queue from the first row with
Started = 0
.
For each row, the sample-specific data are transferred to the method of
the current determination and the parameter
Started
is set to
1
for this
row.
If processing the sample queue is interrupted with
then it can
be continued again with
(with a running chromatogram only
when data acquisition is finished).