2.7 Precolumns and separating columns

761 Compact IC

31

2.7

Precolumns and separating columns

2.7.1

General information on precolumns

The use of easily exchangeable precolumns protects the separating

columns and prolongs their lifetime. The precolumns available from

Metrohm (see section 6.3.2) are either real precolumns or precolumn

cartridges, which are used together with the 6.2821.040 Cartridge head

or the 6.2828.010 Precolumn cartridge holder.

New IC precolumns are normally filled with solution and sealed at both

ends. Before the precolumn is installed in the system, it must be

ensured that this solution is freely miscible with the eluent used

(check manufacturer's specifications).

2.7.2

Precolumns with cartridge head

The 6.1005.020, 6.1005.040, 6.1005.050, 6.1007.010 and 6.1010.010

Precolumn cartridges are installed in the 6.2821.040 Cartridge head as

follows (see Fig. 12):

1 Prepare separating column

Remove end caps from separating column 81.

Unscrew fastening screw from column inlet.

Take steel connector 74 for ferrule out of fastening screw.

2 Install cartridge

Remove end caps for precolumn cartridge 77 (the steel mesh

76 and gaskets 75 are already installed in the cartridge).

Mount steel spacer 78 on Manufit pressure screw 80 (the

steel mesh 79 and gaskets 75 are already installed in the

pressure screw).

Mount precolumn cartridge 77 on the steel spacer 78 (com-

ply with flow direction if specified on the precolumn).

Screw Manufit pressure screw 80 firmly to separating column

81.

Mount steel connector 74 for ferrule on the inlet side of the

precolumn cartridge 77.

Screw on Manufit housing 73 with Manufit pressure screw 80.