4 Operation
761 Compact IC
114
Stop single peak mode
Disable peak end detection. All peaks of any
group after this event will be treated as one
peak. The group end will be treated as peak
end.
Start single peak mode
Set normal detection mode, when valley causes
perpendicular drop or skim line separation.
Split peak
Terminates the current peak and starts a new
one.
Integration events for baseline
Enable valley-to-valley
Disable valley-to-valley
Enable valley-to-valley
Disable perpendicular drop peak separation. All
peaks are considered to be baseline-separa-
ted. The bottom of the valley becomes the
baseline point.
Disable valley-to-valley
Enable perpendicular drop peak separation
(default setting).
Set horizontal
Set back
Set normal
baseline horizontal base
baseline
Set horizontal baseline
Set horizontal baseline for all peaks except the
last one of adjacent peaks that are not sepa-
rated. The baseline is drawn forwards from the
peak start point.
Set back horizontal base
Set horizontal baseline for all peaks except the
first one of adjacent peaks that are not sepa-
rated. The baseline is drawn backwards from
the peak end point.
Set normal baseline
Set default baseline detection mode.