2 Installation

761 Compact IC

28

3 Mount CO

2

absorber tube

First place a piece of cotton wool 70 followed by CO

2

ab-

sorber 69 (e.g. Merck soda-lime pellets with indicator, no.

6839.1000) in the large opening of absorber tube 72 and then

close it with the plastic lid.

Fasten absorber tube 72 to bottle attachment 66 with the aid

of SGJ clip 71.

4 Connect aspirating tubing to pump

Insert the free end of aspirating tubing 63 into one of the

openings 12 in the interior of the 761 Compact IC (see Fig. 2).

Pull aspirating tubing 63 sufficiently far into the interior of the

761 Compact IC, cut off to the required length and push at

least 5 mm of it onto aspirating capillary 40 (see Fig. 3 and

Fig. 4) of the high-pressure pump (it may be necessary to use

emery paper).

If necessary, fix aspirating tubing 63 in the required position

in the interior with the aid of the Y.107.0150 self-adhesive

strap.

Instead of threaded stopper

65

, the 4.420.0311 Tubing nipple (M6)

which belongs to the 6.1602.160 Bottle attachment accessories can

be used together with the included second E.301.0021 O-ring for

recycling operation (see section 5.2.3), in which the outlet capillary

from the detector block is led back into eluent bottle

67

.

63
64
65

66

72

68

71

70

69

67

Fig. 11: Connection of eluent bottle

63 Aspirating tubing (6.1834.010)
64 Tubing nipple (4.420.4300; M8)

with E.301.0021 O-ring

65 Threaded stopper (6.1446.040; M6)
66 Bottle attachment (6.1602.105)
67 Eluent bottle (6.1608.070)
68 Aspirating filter (6.2821.090)
69 CO

2

absorber

70 Cotton wool
71 SGJ clip (6.2023.020)
72 Absorber tube (6.1609.000)