2.6 High-pressure pump

761 Compact IC

27

2.6.3

Connecting the eluent bottle

The eluent supply line from the storage bottle to the high-pressure

pump is connected as follows (see Fig. 11):

Only

degassed (with N

2

, He or vacuum) and

microfiltered (0.45

µ

m

filter)

eluents should be used!

The 6.1608.070 Eluent bottle (2 L) supplied is

not suitable for

vacuum degassing. Use a pressure-resistant container for this.

Care must be taken that the

eluent used is freely miscible with any

solvent remaining in the pump head (the pump head leaves the

factory filled with either isopropanol or methanol/water). If this is not

the case then the pump must first be rinsed with a solvent which is

miscible with both the previous eluent and the following eluent (e.g.

acetone).

1 Prepare eluent bottle

Prepare, microfilter (0.45

µ

m microfilter) and degas (with N

2

,

He, or vacuum) the suitable eluent for the required application

and separating column.

Fill eluent into eluent vessel 67 (clear glass, 2 L).

Place eluent bottle 67 at the front in bottle holder 10 on the

761 Compact IC (see Fig. 1).

2 Install bottle attachment

Firmly screw threaded stopper 65 (6.1446.040; part of

6.1602.160) into the smaller threaded opening (M6) of bottle

attachment 66 (6.1602.105; part of 6.1602.160).

Firmly screw aspirating filter 68 onto aspirating tubing 63.

Pull the other end of aspirating tubing 63 through the larger

threaded opening (M8) of bottle attachment 66 from below.

Push O-ring (E.301.0021; part of 6.1602.160) over the free

end of aspirating tubing 63 and move it towards bottle at-

tachment 66.

Push tubing nipple 64 (4.420.4300; part of 6.1602.160) over

the free end of aspirating tubing 63, move it as far as re-

quired towards bottle attachment 66 and screw it loosely in

the larger opening of bottle attachment 66.

Insert aspirating tubing 63 with screwed-on aspiration filter 68

into eluent bottle 67 and screw bottle attachment 66 onto

eluent bottle 67.

Pull aspirating tubing 63 so far through the opening of tubing

nipple 64 that aspirating filter 68 is touching the bottom of

eluent bottle 67.

Fix aspirating tubing 63 in place by screwing shut tubing

nipple 64.