6 appendix, 1 technical data, 1 conductivity measurement – Metrohm 761 Compact IC User Manual
Page 199: 2 conductivity detector, 1 technical, Data, 1 conductivity, Measurement, Detector, 8 rs232
6.1 Technical data
761 Compact IC
189
6 Appendix
6.1 Technical
data
6.1.1 Conductivity measurement
Measurement range 1
0
…
1000
µ
S/cm (resolution: 0.56 nS/cm)
Measurement range 2
0
…
250
µ
S/cm (resolution: 0.14 nS/cm)
Measurement range 3
0
…
50
µ
S/cm (resolution: 0.028 nS/cm)
Maximum error
±
1 % of full scale value and
±
1 % of measurement value (k = 16.7/cm)
Linearity
Deviations <
±
0.5 % of full scale value
Noise
Measurement range 1
Measurement range 2
Measurement range 3
typ. 10 nS/cm
typ. 2.5 nS/cm
typ. 0.5 nS/cm
Drift (electronic)
typ. < 10 ppm/h of full scale value
Temperature dependence
typ. < 40 ppm/°C of full scale value
Reserve range
> 33 % (k = 16.7/cm)
Sampling rate
10 measurements/s (fixed)
6.1.2 Conductivity detector
Construction
Thermostatted conductivity detector with
2 ring-shaped steel electrodes
Measurement principle
Alternating current measurement with 1 kHz
frequency and ca. 1.7 V amplitude (peak to
peak).
Effective cell volume
0.8
µ
L
Cell constant
approx. 17 /cm (the exact value is printed on the
detector)
Maximum back pressure
for measuring cell
5.0 MPa (50 bar)
Thermostatting
Connectable dynamic control to adjustable
operating temperature
Operating temperature
Adjustable in steps of 5°C from 25
…
45°C
Max. temperature deviation
±
2.5°C
Heating time
≥
30 min
Temperature stability
≤
0.01°C at constant ambient temperature
Connection for detector block Dsub 15 pin (female)