Metrohm 761 Compact IC User Manual
Page 150
4 Operation
761 Compact IC
140
?
Some points of corresponding peak are out of
ADC or detector range – suspicious result.
!
Component concentration is outside of mini-
mum (
min C
) and maximum limits (
max C
), set in
the component table.
p
Special component (it has one of the special
flags checked on in the
COMPONENT
window).
N
The component response is outside of cali-
brated region.
In the case of peaks additional letters precede
the component type information, indicating,
how peaks are separated from other peaks, for
example:
BD_
Peak that starts on the baseline (
B
) and ends
on the drop line (
D
) that separates it from an-
other adjacent peak.
BBR
Special case of a rider (
R
) peak that is tangen-
tially separated from the main one. A main peak
will have a third letter
H
(horse) in this case.
The complete component type may look like:
BBD : !R
.
group
Number of the group for the component.
spectral ratio
Not used for 761 Compact IC.
name
Name of the component.
file name
File name of chromatogram. This column is
very useful for the purpose of processing of ex-
ported data.
ident
Title (identifier) for the chromatogram. This col-
umn is useful for the processing of exported
data.