5 Notes – Maintenance – Faults

761 Compact IC

186

5.3

Faults and malfunctions

5.3.1 Error

messages

If any type of malfunction occurs during operation of the 761 Compact

IC, this is shown by error messages in the PC program, which appear

either in an error window or in the

SYSTEM STATE

window.

Follow the instructions given in the error window and close this win-

dow with

.

You will find further details of the error messages of the

SYSTEM STATE

window, their possible causes and the procedure for their rectification in

section 4.3.8.

5.3.2

Malfunctions and their rectification

If difficulties appear with the 761 Compact IC during analyses, their

causes are best investigated in the order separating column

high-

pressure pump

eluent

connections. Several of the malfunc-

tions which may appear are listed in the following table with details of

possible causes and countermeasures.

Malfunction

Cause

Rectification

Baseline with high
noise level,
pulsation

Contaminated pump
valves

Faulty piston seals

Clean the valves (see
section 5.1.5)

Replace the piston seals
(see section 5.1.5)

Drift of the
baseline

Thermal equilibrium not
yet reached

Leak in system

Evaporation of organic
solvent in eluent

Condition system with
heating switched on

Check connections and
make leakproof

Ensure better closure of
eluent supply vessel

Considerable
pressure drop

Leak in system

Check connections and
make leakproof

Considerable
pressure rise

Contamination of the
filter in the 6.2821.120
Filter unit PEEK

Contamination of the
column inlet filter

Change of column
packing by injection of
contaminated samples

Replace the 6.2821.130
Filter (see section 2.3.6)

Clean or replace
6.2821.020 Steel mesh(es)

Regenerate column (see
section 5.1.1) or replace
column

Note:
Samples should always
be microfiltered.