Metrohm 761 Compact IC User Manual

Page 67

background image

3.3 Calibration

761 Compact IC

57

9 Enter information for determination

Click on

or select

Passport

of the

Method

menu. The

passport opens with the

General

page, which contains the

general information about the determination.

Enter a title for the calibration as

Ident

(e.g.

Standard

).

Change the analysis time under

Duration

if required.

Change other parameters on the passport tabs to character-

ize sample and acquirement conditions if required.

10 Switch injection valve to "FILL" position

Click on the

button of the 761 icon in the system

window to switch the injection valve to the "FILL" position. At

the same time, the suppressor module is switched to the next

position.