3.3 Calibration
761 Compact IC
59
⇒
Click on
. The
Integration
parameters
window is
opened and the chromatogram is reintegrated.
⇒
Repeat this procedure for all other integration parameters
until the result satisfies your expectations. Click on
for each parameter change.
⇒
Close the
Integration
parameters
window with
.
15 Modify peak allocation
⇒
Click on
or select
Calibration / Components
from the
Method
menu. The prepared table for the components fluo-
ride, chloride, nitrate, and sulfate is displayed below the
chromatogram. For all peaks which could be allocated un-
ambiguously to a component this table contains the peak
number in the
Peak
column and the retention time in the
Time
column. In the chromatogram itself this retention time is indi-
cated by the cursor (vertical line).
⇒
If the
Peak
column contains a
0
the peak must be manually
allocated to a component. In this case, insert the peak num-
ber from the chromatogram into the
Peak
column and click
the
Time
field on this row with the mouse. The corresponding
retention time is then inserted automatically as the new time
for this component.
⇒
For each correctly allocated peak click on the retention time
in the
Time
column. This time is automatically shown in the
chromatogram by the cursor. If the retention time is not lo-
cated in the area of the middle of the peak then the value in
the
Time
column should be adapted accordingly.