2.9 Putting into operation

761 Compact IC

47

parameters.

3 Start system

Make sure, that the aspirating tubing 63 for the high-pressure

pump is immersed in the eluent.

Select

Startup determination

from the

Control

menu in the

system window. The high-pressure pump and the peristaltic

pump are started, at the same time, a chromatogram window

is opened where the baseline is recorded continuously. The

suppressor module is automatically switched to the next posi-

tion every 20 min and conditioned in this way.

4 Set contact pressure for pump tubings

Press contact pressure lever 49 on both tubing cartridges 48

upwards until regeneration and rinsing solution just start to be

drawn in.

Then press contact pressure lever 49 upwards until it clicks

once more to obtain optimal contact pressure.

5 Check for leaks

Check all capillaries and their connections between the high-

pressure pump and the detector block for escaping liquid. If

eluent escapes anywhere, the appropriate compression fitting

must be tightened further or changed.