Metrohm 761 Compact IC User Manual
Page 28
2 Installation
761 Compact IC
18
2.3.6
Filter unit PEEK
One 6.2821.120 Filter unit PEEK (see Fig. 8) is already installed be-
tween the high-pressure pump and the injection valve at the 761 Com-
pact IC. This filter unit serves to avoid contamination by abrasive parti-
cles of the piston seals.
The two other filter units PEEK supplied with the 2.761.0020 Compact
IC (with suppressor) are installed between the pump tubings of the
peristaltic pump and the inlet capillaries for regeneration and rinsing so-
lution (see section 2.8.2). These filter units serve to protect the suppres-
sor module from foreign particles and bacterial growth.
The Filter unit PEEK 36 consists of the filter-housing 59, the filter-screw
57 and the 6.2821.130 filter 58. For the connection of capillaries 56
PEEK compression fittings 54 (6.2744.010 or 6.2744.070) must be
used. New filters 58 are available as an option with the ordering number
6.2821.130 (set of 10).
For the connection of the filter unit, please note the flow direction
arrow printed on the housing.
56
54
57
58
59
54
56
36
Fig. 8: 6.2821.120 Filter unit PEEK
36 Filter unit PEEK (6.2821.120)
57 Filter-Screw of Filter Unit
Part of 6.2821.120 Filter unit
54 Compression fitting (6.2744.010)
58 Filter 6.2821.130
Part of 6.2821.120 Filter unit
56 Capillary
6.1831.010 PEEK capillary or
6.1822.010 PTFE microcapillary
59
Filter-Housing of Filter Unit
Part of 6.2821.120 Filter unit