4.3 Systems
761 Compact IC
93
Event
Events for automatic remote signal output:
Inject
Switching of the injection valve to the "INJECT"
position.
Fill
Switching of the injection valve to the "FILL" po-
sition.
Suppressor step
Switching of the suppressor module to the next
position.
Pump stop (P max/min)
Pump stopped because pressure limits are ex-
ceeded.
Hardware error
Hardware error detected at the 761 Compact IC
(high-pressure pump, injection valve, or sup-
pressor not working correctly).
Alarm leak detector
Leak detector has detected solvent in the in-
struments interior.
Remote line
Set remote output lines 1
…
8.
Selection:
0
(line off, inactive, open)
1
(line on, active, 0 V)
p
(pulse output)
*
(leave line in current status)
PowerOn values
Display of the power on startup values for re-
mote output lines set on the
Hardware
tab.
Pulse length
Length of a pulse in ms.
Alarm stops
The
Alarm stops
tab of the hardware settings window defines the events
for which the 761 Compact IC is stopped immediately. At an alarm
stop, high-pressure pump and peristaltic pump are stopped immedi-
ately, the running determination and the active sample queue are also
stopped.
Event
Events for alarm stop:
Pump stop
Pump stopped because pressure limits are ex-
ceeded.
Hardware error
Hardware error detected at the 761 Compact IC
(high-pressure pump, injection valve, or sup-
pressor not working correctly).