4.3 Systems

761 Compact IC

93

Event

Events for automatic remote signal output:

Inject

Switching of the injection valve to the "INJECT"

position.

Fill

Switching of the injection valve to the "FILL" po-

sition.

Suppressor step

Switching of the suppressor module to the next

position.

Pump stop (P max/min)

Pump stopped because pressure limits are ex-

ceeded.

Hardware error

Hardware error detected at the 761 Compact IC

(high-pressure pump, injection valve, or sup-

pressor not working correctly).

Alarm leak detector

Leak detector has detected solvent in the in-

struments interior.

Remote line

Set remote output lines 1

8.

Selection:

0

(line off, inactive, open)

1

(line on, active, 0 V)

p

(pulse output)

*

(leave line in current status)

PowerOn values

Display of the power on startup values for re-

mote output lines set on the

Hardware

tab.

Pulse length

Length of a pulse in ms.


Alarm stops
The

Alarm stops

tab of the hardware settings window defines the events

for which the 761 Compact IC is stopped immediately. At an alarm

stop, high-pressure pump and peristaltic pump are stopped immedi-

ately, the running determination and the active sample queue are also

stopped.

Event

Events for alarm stop:

Pump stop

Pump stopped because pressure limits are ex-

ceeded.

Hardware error

Hardware error detected at the 761 Compact IC

(high-pressure pump, injection valve, or sup-

pressor not working correctly).