3 operating tutorial, 1 requirements, 2 preparations – Metrohm 761 Compact IC User Manual

Page 63: 3 calibration, 4 sample, Determination

background image

3.1 Requirements

761 Compact IC

53

3 Operating

tutorial


This section introduces you to the operation of the 761 Compact IC by

means of a brief operating tutorial which describes the basic operat-

ing steps needed for the recording of an ion chromatogram using one

of the system files supplied.
The determination of the anionic content of a drinking water sample

with the METROSEP Anion Dual 2 column with chemical suppression

is used an illustrative example. Please note that the steps and pa-

rameter settings described apply only to this example. If you use a

different separating column and system file, the procedures described

in the tutorial must be modified appropriately.
For further explanations of the operation, please refer to section 4.





3.1 Requirements

For the determination of anions in drinking water described in this tuto-

rial, the following instruments, accessories and solutions are required:

2.761.0020 Compact IC

with suppressor module

6.1825.210 Sample loop (20

µ

L, PEEK)

already integrated in the 2.761.0020 Compact IC

6.1006.100 IC anion column METROSEP Anion Dual 2

Eluent

2 mmol/L NaHCO

3

/ 1.3 mmol/L Na

2

CO

3

in dist. H

2

O

flow: 0.8 mL/min

Standard

Standard solution with 0.5 mg/L F

, 5 mg/L Cl

and

10 mg/L each of NO

3

and SO

4

2–

(in dist. H

2

O)