Metrohm 761 Compact IC User Manual

Page 42

background image

2 Installation

761 Compact IC

32

74

73

75 76 77 76 75 78 75

54

28

79 80

81

Fig. 12: Installation of precolumn cartridges with cartridge head

28 Column connection capillary from

injector

77

Precolumn cartridge

54

Compression fitting (6.2744.010)

78

Steel spacer (6.2821.080)

73

Manufit housing

79

4 Steel meshes (6.2821.020)

74

Steel connector

(of IC separating column)

80 Manufit pressure screw

75 PTFE gasket (6.2821.010)

81 IC separating column

76

2 Steel meshes (6.2821.020)


3 Connect precolumn

Fit PEEK compression fitting 54 to the column connection

capillary 28 mounted on the injection valve (see section

2.3.5).

Screw column connection capillary 28 to Manufit housing 73.


The precolumn cartridge built into the cartridge head can only be

rinsed together with the separating column (see section 2.7.7/8). This

increases the rinsing time by approx. 20 min.