Flow correction
Factor for correction of the difference between
displayed and actual flow rate of the high-
pressure pump.
Range:
0.9
...
1.09
The flow correction is determined by measure-
ment of the actual flow rate with the aid of a
measuring cylinder as follows:
Flow correction
=
Displayed flow rate
Measured flow rate
Cell constant
Cell constant of the conductivity cell for correct
display of the absolute conductivity. Enter the
cell constant printed on the detector block into
this field.
Range:
0.1
...
1000 /cm
For a precise determination of the cell constant,
pump a calibration solution of known conduc-
tivity through the IC system, observe the dis-
played conductivity and change the cell con-
stant until the correct conductivity value is
displayed.
Possibility for changing the break times for in-
jection valve
Valve
and suppressor module
Sup-
pressor
.
These values should only be changed after
consulting the Metrohm Service.
Outputs
The
Outputs
tab of the hardware settings window defines remote output
signals to be set automatically if specific events occur.