Metrohm 761 Compact IC User Manual

Page 170

background image

4 Operation

761 Compact IC

160

No

Row

number.

System

System file to be used for the determination.

The system is automatically opened and con-

nected.

Ident

User defined identifier for chromatogram. Will

be placed into appropriate passport field when

starting the chromatogram.

Vial

Autosampler vial position to take sample from.

This value will not be transferred automatically

to the autosampler. Make sure that the vial

positions entered in this column are identical

to the vial positions used at the autosampler.

Volume

Injected volume of sample in

µ

L.

Dilution

Sample dilution prior to injection.

Amount

Sample amount. If this value is different for the

calibration run (c) and the sample run (s), the

component concentrations of the sample are

calculated as follows:

C

s

=

C

c

Amount

s

/

Amount

c

Internal standard amount

Concentration of the internal standard compo-

nent for relative concentration calculations.

Level

Calibration level (see section 4.4.5) for the

sample: Level

0

stays for normal analysis, levels

1

and greater for calibration runs. Correct filling

of this column enables to calculate calibration

coefficients automatically.

Injections

Number of injections for the same vial position

(only 1 injection is permissible for

Level

1

).

This value will not be transferred automatically

to the autosampler. Make sure that the

numbers of injections entered in this column

are identical to the numbers of injections used

at the autosampler.

Done

Indication whether the sample determination

has been done or not:

0

determination not yet done

1

n

determination or injection done

Sample Info 1

First sample description.

Sample Info 2

Second sample description.