2 connection of the 766 ic sample processor – Metrohm 761 Compact IC User Manual

Page 60

background image

2 Installation

761 Compact IC

50

4 Settings at the 750 Autosampler

Set parameter "

Run time

" to

0.0

to activate the external start

of the 750 Autosampler (see section 3.5 of the 750 Instruc-

tions for Use). Filling the sample loop of the injection valve

only takes place when the impulse produced by the 761

Compact IC at remote output 3 (see point 5) makes contact

at the "EXTERNAL INJECT INPUT" of the 750 Autosampler.

The run time will then be determined externally.

5 Settings in the «761 Compact IC» program

A time program must be drawn up for the selected system

which first switches the injection valve to the "Fill" position,

then produces an impulse at remote lead 3 to start sample

addition by the 750 Autosampler and finally switches the in-

jection valve to the "Inject" position. The following example

shows such a program:



2.10.2 Connection of the 766 IC Sample Processor

The 766 IC Sample Processor available from Metrohm as an option is

an automatic sampler for ion chromatography. It accommodates max.

127 samples each of 2.5 mL or 11 mL, which are automatically trans-

ferred to the sample loop attached to the injection valve of the 761

Compact IC. Sample changing and filling the sample loop are started

by a remote signal produced by the 761 Compact IC (761 Compact IC

as "Master").
The 6.2141.110 Cable is required for connecting the 766 IC Sample

Processor. The 766 IC Sample Processor is connected to the 761

Compact IC as follows:

1 Electrical connection 761 – 766

Connect the end of the 6.2141.110 Cable marked with "766"

to the remote interface 22 of the 766 IC Sample Processor

(see 766 Instructions for Use).