beautypg.com

3 eluents, 4 peristaltic pump – Metrohm 761 Compact IC User Manual

Page 181

background image

5.1 Practical notes on ion chromatography

761 Compact IC

171

5.1.3 Eluents

Treatment
For the preparation of the eluents only chemicals of a purity degree of

at least "p.a." should be used. For diluting please use only high purity

water.
Fresh eluents should always be microfiltered (0.45

µ

m filter) and de-

gassed (with N

2

, He or vacuum). For alkaline eluents and eluents with

low buffering capacity one should preferably use a CO

2

absorber (see

section 2.6.3).
The supply vessel containing the eluent must be closed as tightly as

possible to avoid excessive evaporation. This is primarily important with

eluents containing organic solvents (e.g. acetone), the evaporation of

which can lead to drifts in the long term. If work is performed in a very

sensitive range, even if one drop of condensate falls back in the eluent

this can cause a noticeable change in the background conductivity.

Influence of various parameters on anion columns

Concentration:

An increase in the concentration usually leads

to shorter retention times and quicker

separation, but also to a higher background

conductivity.

pH:

pH alterations lead to shifts in the dissociation

equilibrium and thus to changes in the

retention times.

Organic modifiers:

Addition of an organic solvent (e.g. methanol,

acetone, acetonitrile) to aqueous eluents

generally accelerates lipophilic ions.

Eluent change
When the eluent is changed, it must be ensured that no precipitates

can be formed. Solutions used in direct succession must therefore be

miscible. If the system has to be rinsed with an organic solution, several

solvents with increasing or decreasing lipophilic character may possibly

have to be used (e.g. water

acetone

chloroform).

5.1.4 Peristaltic

pump

The pump tubings used by the peristaltic pump are consumable mate-

rial with a limited lifetime and should be exchanged at regular intervals

(approx. every 4 weeks under continuous use; see section 5.2.10).
The working life of pump tubing depends to a considerable extent on

the contact pressure. This is why the contact pressure must be correctly

set as described in section 2.9.2. If the pump is to remain switched off

for a lengthy period of time the tubing cartridges 48 should be raised

completely by loosening the snap-action lever 51 on the right-hand side

(the set contact pressure remains unchanged).