2 Installation

761 Compact IC

42

pump tubing 92.

Mount a 6.2744.010 Compression fitting at the end of the

aspirating tubing 90 and screw this compression fitting on to

the coupling 91 (see Fig. 16).

3 Prepare supply bottle for H

2

O

Prepare rinsing solution suited for the desired application and

separating column (normally dist. H

2

O).

Fill rinsing solution into supply bottle 103 (amber glass, 1 L)

and label the bottle.

Screw bottle attachment 102 on to supply bottle 103.

Place supply bottle 103 beside the other supply bottle on the

bottle rack 10 on the 761 Compact IC (see Fig. 1).

4 Connect aspirating tubing for H

2

O

Prepare aspirating tubing 89: Cut a piece of the 6.1803.020

PTFE tubing to the required length (normally ca. 120 cm).

Pull one end of aspirating tubing 89 through a PEEK com-

pression fitting 54 (6.2744.010) so that approx. 30 cm of the

tubing projects.

Screw PEEK compression fitting 54 with tubing into one

opening of bottle attachment 102 attached to the rinsing so-

lution storage bottle and tighten it so that the tubing is firmly

held.

Insert the free end of aspirating tubing 89 into one of the

openings in 12 of the 761 Compact IC (see Fig. 2) from

above and pull it sufficiently far into the interior.

Mount a coupling 91 (6.2744.030) to the inlet end of the front

pump tubing 93.

Mount a 6.2744.010 Compression fitting at the end of the

aspirating tubing 89 and screw this compression fitting on to

the coupling 91 (see Fig. 16).

89/90

54
102
103

Fig. 17: Connection of supply bottles

54 PEEK compression fitting

(6.2744.010)

89 Aspirating tubing for H

2

O

6.1803.020 PTFE tubing

90 Aspirating tubing for H

2

SO

4

6.1803.020 PTFE tubing

102 Bottle attachment (6.1602.150)

103 Supply bottle (6.1608.023)