2.7.8
Installation of the separating column with suppressor
With the 2.761.0020 Compact IC with suppressor module, the IC sepa-
rating column is first connected to the injector or precolumn. The con-
nection to the suppressor module and the detector block is described
in section 2.8.
1 Connect column to injector
•
Remove end caps from column 81.
•
Screw inlet end of separating column 81 (note flow direction)
to column connection capillary 28 or to the already installed
precolumn (procedure see section 2.7.7).
2 Rinse column
•
Place a beaker beneath the column outlet.
•
Open software window for manual system control.
•
If necessary, modify
Flow rate
to the value suited for the
inserted separating column and click on
to
send this value to the 761 Compact IC.
•
Switch on high-pressure pump (
IC pump
) by clicking
and rinse column with eluent for ca. 10 min.
•
Switch off high-pressure pump by clicking
.
3 Fix column
•
Insert one or two column holders 88 (6.2027.030, 6.2027.040
or 6.2027.050) in the mounting rails 27 and fasten separating
column 81 in the column holder 88.
27 Mounting rail
for column holder 88
28 Column connection
capillary
Connection injection
valve 32 –
separating column 81
32
Injection valve
45
Inlet capillary for
detector block
PEEK capillary, fixed
mounting
46 Detector block
(1.733.0110)
81 Separating column
88 Column holder
(6.2027.0X0)