4 Operation

761 Compact IC

116

Force horizontal base back

Set horizontal baseline for a single peak. The

baseline is drawn backwards from the peak

end point. The intersection of the baseline with

the signal is defined as peak start point.

Cancel horizontal base back

Clear

Force horizontal base back

mode.



Enable baseline

Disable baseline

penetration

penetration

Enable baseline penetration

Allow crossing of the signal by the baseline.

Disable baseline penetration

Clear

Enable baseline penetration

mode. The

baseline is aligned to the bottom value of the

signal.

Integration events for integration parameters

Set width

Sets

new

Width

parameter that supersedes de-

fault linear growth of expected peak.

Set slope

Sets

new

Slope

value.

Set min height

Sets

new

Minimum peak height

value.

Set rider ratio

Sets new value of

Rider ratio

parameter.