2 Installation

761 Compact IC

16

Insert the cable permanently attached to the detector block

46 in one of the openings 11 and the outlet capillary in one of

the openings 13 of the rear panel 17.

Replace rear panel 17 and screw to the 761 Compact IC

using the four knurled screws 15.

3 Connect detector block

Plug the gray connecting cable permanently attached to the

detector block 46 into connection 23 "Detector Block" of the

761 Compact IC and fasten to the instrument by tightening

the screws in the cable connector (see Fig. 2).

4 Connect waste container

Lead the outlet capillary of the detector block 46 to a suffi-

ciently large waste container and fix in place.

2.3.2

Connection of syringe and aspirating tubing

For manual filling of the sample loop 29 mounted on the injection valve,

the 6.2816.020 Syringe and the PTFE aspirating tubing 4 already

screwed to the valve are needed. These accessories are mounted or

adjusted as follows:

1 Connect syringe

Push 6.2816.020 Syringe (without needle) as far as it will go

into connection socket 2 (see Fig. 1).

2 Adjust aspirating tubing

Loosen the rotary nipple 31 screwed onto the interior side of

feedthrough 3 (see Fig. 3 and Fig. 4).

Pull PTFE aspirating tubing 4 (see Fig. 3 and Fig. 4) by hand

out of feedthrough 3 as far as desired.

Retighten rotary nipple 31 on the interior side of feedthrough

3 to fix the aspirating tubing 4 in place.

2.3.3

Connection of the drain tube for the inner compartment

The 761 Compact IC has a connection at the front to which a drain tube

for discharged liquids in the inner compartment can be attached. Pro-

ceed as follows:

1 Connect drain tube

Mount 6.1816.020 Silicone tubing on connection nipple 6

(see Fig. 1).

2 Lead drain tube to collecting vessel

Lead the other end of the drain tube to a suitable collecting

vessel and fix in place.