2.9 Putting into operation

761 Compact IC

45

2.9

Putting into operation

2.9.1

Putting into operation without suppressor

Before sample solutions can be injected at the 2.761.0010 Compact

IC (without suppressor), the entire system must be tested for leaks and

then conditioned with eluent until the baseline is stable. Proceed as fol-

lows:

1 Open and connect system

Start the «761 Compact IC» PC program, if it has not already

been started (see section 2.5.3).

Select

File / Open / System

in the main window. Select the

system file suited for the inserted separating column (e.g.

n-

01.smt

) and click on

.

Select the

Connect to workplace

item from the

Control

menu of

the system window.

2 Open control window

Double-click the 761 icon in the system window. The control

window for manual control of the 761 Compact IC appears,

which indicates conductivity, pressure and current system

parameters.