Metrohm 761 Compact IC User Manual
Page 34
2 Installation
761 Compact IC
24
•
Select the
Connect to workplace
menu item of the
Control
menu
in this window.
•
If the connection between PC and Compact IC is working the
message
Hardware settings file for 761 unit with serial number
'#####' not found! Create?
will appear. Click
in order to
create the configuration file
#####.761
for this instrument.
•
If the connection between PC and Compact IC does not work
then the message
Detection of hardware failed[761 Compact IC
[ COM# ]]
appears in the
SYSTEM STATE
window. In this case
check whether the instrument has been switched on, whether
the connection cable is connected up properly and whether
the COM interface has been set correctly (see point 2). Then
repeat point 3.
4 Hardware settings
Of the general hardware settings only the input of the cell constant
is described. Standard settings can normally be used for all other
parameters.
•
Click the 761 icon using the right mouse button and select
the
Hardware
item. The
Hardware settings
window is opened:
•
In the
Cell constant
field enter the cell constant which is printed
on the 1.732.0110 Detector block (see section 2.3.1).
•
Click on
to close the window and save the settings.