2 Installation

761 Compact IC

34

85

84

83

86

87

54

28

83

82

82

54

56

Fig. 13: Installation of precolumn glass cartridges with cartridge holder

28 Column connection capillary from

injector

84

Sleeve for precolumn cartridge

54

Compression fitting (6.2744.010)

85

Precolumn cartridge (6.1006.030)

56

PEEK capillary (6.1831.010)

86

Connection piece

for connection precolumn – column

82

End fitting

87 Screw cap for column

83

Screw cap for precolumn

2.7.4

IC anion precolumn SUPERSEP

The 6.1009.010 IC Anion Precolumn SUPERSEP has two connections

for PEEK capillaries and is installed as follows:

1 Connect precolumn

Remove end caps from the precolumn.

Fit PEEK compression fitting 54 to column connection capil-

lary 28 mounted on the injection valve (see section 2.3.5).

Screw precolumn to column connection capillary 28.

Cut a piece from the 6.1831.010 PEEK capillary as short as

possible and fit with PEEK compression fittings 54 (see sec-

tion 2.3.5).

Fasten the prepared capillary to the other end of the precol-

umn.

2 Rinse precolumn

Place a beaker beneath the outlet capillary of the precolumn.

Open software window for manual system control.

If necessary, modify

Flow rate

to the value suited for the

inserted separating column and click on

to

send this value to the 761 Compact IC.

Switch on high-pressure pump (

IC pump

) by clicking

and rinse precolumn with eluent for ca. 10 min.

Switch off high-pressure pump by clicking

.