4.5 Chromatograms
761 Compact IC
149
Other graphical functions
761 COMPACT IC / View / X full scale
By selecting this menu item or pressing [ Ctrl ] + [ Home ] the X axis is
scaled so that the chromatogram perfectly fits the window horizontally.
761 COMPACT IC / View / Y full scale
By selecting this menu item or pressing [ Ctrl ] + [ End ] the Y axis is
scaled so that the chromatogram perfectly fits the window vertically.
761 COMPACT IC / View / View all
By selecting this menu item or pressing [ Alt ] + [ V ] the X and Y axes
are scaled so that the chromatogram perfectly fits the window horizon-
tally and vertically.
761 COMPACT IC / View / Recorder autoscale
This option allows to look at the chromatogram so that the last point is
always visible on the screen while data are acquired.
If the
Recorder autoscale
option is switched on, the following autoscaling
rules apply:
– If the last acquired point comes out of the window down, autozero
is performed.
– If the last point is too high, the recorder scale shrinks twice until
the point comes to screen.
– If the last point comes outside of the plotting area to the right, the
window is shifted half-screen right.
If the
Recorder autoscale
option is switched off, the window scale does