Metrohm tiBase 1.0 Manual User Manual
Page 94

3.2 Administration
■■■■■■■■■■■■■■■■■■■■■■
86
■■■■■■■■
tiBase 1.0
Selection
on | off
Default value
off
FDA default: off
Automatic logout after
If this check box is activated then the user will be logged out automatically
when no operating functions have been carried out with the keyboard or
the mouse within this time. After this automatic logout, only a member of
the same user group as the one to which the previously logged-off user
belongs or a user with administrator rights can log in again.
Selection
on | off
Default value
off
FDA default: off
Range
1 ... 60 min
Default value
10 min
FDA default: 10 min
Note
The immediate termination of all determinations on all activated work-
places with the emergency stop button still remains possible after an
automatic logout.
Password policies
Enforce unique password
If this check box is activated then this ensures that a user can use a given
password only once.
Selection
on | off
Default value
off
FDA default: on
Enforce use of special characters
If this check box is activated then this ensures that the password must
contain at least one special character (@, #, ~ etc.).
Selection
on | off
Default value
off
FDA default: off
Minimum password length
If this check box is activated then this ensures that the password must
contain the indicated number of characters.