5.2 Determinations
■■■■■■■■■■■■■■■■■■■■■■
300
■■■■■■■■
tiBase 1.0
7 In the dialog window Reprocessing click on [OK].
Each determination that has been modified by reprocessing will be
saved as a new version with a version number increased by +1 and the
dialog window Reprocessing will be closed. This button is inactive
when the recalculation has not yet been triggered and if not all the
selected determinations could be reprocessed.
Modify statistics
Note
The tab Statistics will only be shown when the last determination (and
only this one) is selected from a set of determinations which, because of
the statistics defined in the method, belong together.
1 Change to the Statistics tab.
2 In the field Result name, select the result whose Result value is to
be displayed.
3 Select desired determination whose result value(s) should be switched
on or off for the statistics.
4 If only the selected result of the determination has to be switched on
or off for the statistics, click on [Result on/off].
If the result is switched off, an asterisk (*) appears behind the result
value, is it switched on again, the asterisk disappears.
5 If all results of the selected determination have to be switched on or
off for the statistics, click on [Determination on/off].
If the determination is switched off then an asterisk (*) appears behind
all result values in the table and the line is shown as inactive (gray); if
it is switched on again then the asterisks will disappear.