2 General program functions
tiBase 1.0
■■■■■■■■
27
Variable
name
(PC-Control)
Variable
name
( tiBase 1.0 )
Description
Command
.RD#
.RE{x}.[DRT]
Mean dosing rate in window x
(1…9) in mL/min
STAT
.MCM
.RES
Result calculated from the stand-
ard addition curve (concentration
of the selected unit)
STDADD
.MSL
.SLO
Electrode slope of the sensor used
for the command (in % for pH
sensor or mV for ISE sensor, num-
ber) or electrode slope calculated
from the calibration (for CAL
LOOP)
DET pH, DET U, MET pH,
MET U, SET pH, SET U,
STAT, MEAS pH, MEAS U,
MEAS T, MEAS Conc, CAL
LOOP, DOS, STDADD
.MSM
.SME
Start measured value (measured
value after processing the start
conditions) in the unit of the
measured value
DET, MET, SET, KFT, KFC,
STAT
.MST
.STE
Start temperature (temperature
after processing of the start con-
ditions) in °C
DET, MET, SET, KFT, KFC,
STAT
.MTS
.STY
Type of stop with which the com-
mand was stopped: 1 = normal;
0 = manual or after error
DET, MET, SET, KFT, STAT,
MEAS, CAL MEAS,
STDADD, DOS
.MSA
.SVA
Start volume absolute (volume
that was added according to the
start condition "start volume") in
mL
DET, MET, SET, KFT, STAT
.MSP
.SVM
Start volume measured value (vol-
ume which was added according
to the start measured value given
in the start conditions) in mL
DET, MET
.MSS
.SVS
Start volume measured value (vol-
ume that was added according to
the start condition "start slope") in
mL
DET, MET
.MSV
.SVT
Total start volume (volume that
was added in dependence on all
three start conditions) in mL
DET, MET, SET, KFT, STAT
.TITER
.TITER
Titer value of the solution used for
the command (number)
DET, MET, SET, KFT, STAT,
STDADD dos, STDADD
auto, ADD, DOS, LQH
.MVA
.VAR
Variance of the result calculated
from the standard addition curve
STDADD
.MCV
.VOL
Dosed volume
STDADD, ADD, LQH