Determinations - Show calibration curve
Dialog window: Database
▶ Determinations ▶ Show calibration curve... ▶ Calibration
curve
With the menu item Determinations, Show calibration curve... or the
symbol the calibration or standard addition curve is shown for the cali-
bration selected in the dialog window Calibration curve.
Calibration curve for CAL LOOP pH
In the dialog window Calibration curve a tab marked with the command
name showing the calibration curve and calibration data will be displayed
for each CAL LOOP pH command . The command type is shown above the
calibration curve. The curve display shows the measured values and the curve
calculated from these measured values. Listed below are the individual
measured values and the results for Slope, E(0) and Variance.
Calibration curve for CAL LOOP Conc
In the dialog window Calibration curve a tab marked with the command
name showing the calibration curve and calibration data will be displayed
for each CAL LOOP Conc command . The command type is shown above
the calibration curve. The curve display shows the measured values and the
curve calculated from these measured values. Listed below are the individual
measured values and the results for Slope, E(0), c(blank) and Variance.
Calibration curve for CAL LOOP Conc
Standard addition curve for STDADD
In the dialog window Calibration curve a tab marked with the command
name showing the standard addition curve and evaluation data will be dis-
played for each STDADD command . The command type is shown above
the calibration curve. The curve display shows the measured values and the
curve calculated from these measured values. Listed below are the individual
measured values and the results for Determination, Show calibration
curve..., E(0), Determination, Show calibration curve... and Var-
iance.
Note
Calibration curves cannot be printed out directly from the dialog window
Calibration curve. Instead you should use a report with a report tem-
plate which contains the fixed report .