Metrohm tiBase 1.0 Manual User Manual
Page 92

3.2 Administration
■■■■■■■■■■■■■■■■■■■■■■
84
■■■■■■■■
tiBase 1.0
Selection
on | off
Default value
off
FDA default: on
Password monitoring by Windows
If this option is enabled then the password will be monitored according to
the parameters defined in Windows. In this case the parameters for pass-
word protection are deactivated.
Selection
on | off
Default value
off
FDA default: off
Caution
When using password monitoring through Windows, to avoid an unwan-
ted login by the standard defined user Guest in Windows, this must also
either be provided with a password or (preferably) be deactivated.
Note
With Windows 2000, proceed as follows for password monitoring by
Windows:
1.
Run/Execute "secpol.msc".
2.
Under "Local Policies/User Rights Assignment", switch on the guide-
line "Act as part of the operating system" for all desired user groups.
3.
Restart computer.
Maximum number of incorrect login attempts
If this check box is activated then a user will be given the status deacti-
vated as soon as the number of incorrect login attempts defined here has
been exceeded. A user with the status deactivated can no longer start the
program. The counter for incorrect login attempts is reset to zero for all users
at each new start.
Selection
on | off
Default value
off
FDA default: off
Range
2 ... 5
Default value
3
FDA default: 3