beautypg.com

8 common variables, Common variables – Metrohm tiBase 1.0 Manual User Manual

Page 38

background image

2.4 Formula editor

■■■■■■■■■■■■■■■■■■■■■■

30

■■■■■■■■

tiBase 1.0

Variable
name

(PC-Control)

Variable
name

( tiBase 1.0 )

Description

Data type

SNT

.STA

Indication whether the statistics is
activated (1 = yes, 0 = no.

Number

%SC

.STC

Start counter

Number

%SS

Sample table status

1 = Sample table on

0 = Sample table off

Number

2.4.3.8

Common variables

Dialog window: Formula Editor

Variables

Common variables are global variables, which are adopted from the corre-
sponding table of the program part Configuration, where the common
variables can be defined, at the start of the determination and assigned to
the determination. The Variables field of the Formula editor lists all Com-
mon Variables
which are available for the current method, sorted accord-
ing to name.

Syntax
'CV.Variable name.Variable name'

Examples: 'CV.TestDate', 'CV.TestTime.VAL', 'CV.Average-
Temp.UNI'

You can select the common variables directly in the formula editor under
Variables/Common Variables in order to avoid syntax errors.

Available Common Variables

Variable
name

(PC-Control)

Variable
name

( tiBase 1.0 )

Description

Data type

CV01...CV25

[.VAL]

Value of common variables (num-
ber).

Text, number or date/time

Usage
870 KF Titrino Plus

With this instrument the two common variables, CV01 and CV02, are
defined for blind values.

877 Titrino Plus, 848 Titrino Plus, 862 Titrosampler

With these instruments there can be defined up to five method-independent
common variables, CV01 to CV05. Common variables are e.g useful for the
following applications: