2 General program functions

tiBase 1.0

■■■■■■■■

25

Variable
name

(PC-Control)

Variable
name

( tiBase 1.0 )

Description

Command

.ESI#

.EP{x}.MEP

Marking for the endpoint x (1...9);
1 = 1 endpoint, 2 = 2 or more
endpoints

DET, MET, SET, KFT

.EP#

.EP.QTY

Measured value (water) for the
endpoint in mg

KFC

.ET#

.EP{x}.TEM

Temperature for the endpoint x
(1...9) in °C

DET, MET, SET, KFT

.ED#

.EP{x}.TIM

Time for the endpoint x (1...9) in
s

DET, MET, SET, KFT, KFC

.EP#

.EP{x}.VOL

Volume for the endpoint x (1…9)
in mL

DET, MET, SET, KFT

.MCT

.ETE

End temperature (temperature
following processing of the com-
mand) in °C

DET, MET, SET, KFT, KFC,
STAT, MEAS (without T/
Flow), CAL MEAS, STDADD,
DOS

.MCV

.EVT

End volume (total dosed volume
at the end of the command) in mL

DET, MET, SET, KFT, STAT,
DOS

.FM#

.FP{x}.MEA

Measured value for the fixed end-
point x (1…9) in the unit of the
measured value

DET, MET, SET, KFT, STAT

.FP{x}.QTY

Measured value (water) for the
fixed endpoint x (1...9) in mg

KFC

.FT#

.FP{x}.TEM

Temperature for the fixed end-
point x (1…9) in °C

DET, MET, SET, KFT, STAT,
MEAS (without T/Flow)

.FD#

.FP{x}.TIM

Time for fixed endpoint x (1…9)
in s

DET, MET, SET, KFT, KFC,
STAT, MEAS (without T/
Flow)

.FP#

.FP{x}.[VOL]

Volume for the fixed endpoint x
(1...9) in mL

DET, MET, SET, KFT, STAT

.HM#

.HP{x}.MEA

Measured value for HNP x (1...9)
in mV

1)

DET, MET

.HT#

.HP{x}.TEM

Temperature for HNP x (1...9) in
°C

DET, MET

.HD#

.HP{x}.TIM

Time for HNP x (1...9) in s

DET, MET

.HP#

.HP{x}.[VOL]

Volume for HNP (half neutraliza-
tion potential) x (1...9) in mL

DET, MET

#M.MIM

.IME

Initial measured value (measured
value before start conditions are
processed) in the unit of the
measured value (number)

DET, MET, SET, KFT, STAT,
MEAS, CAL MEAS,
STDADD, DOS