Description of errors and corrective actions

Error

code

Error Message

Status marker

shown

Cause

Corrective action

Back end errors

ERR 100 Device reset

1

The device detected an internal

error. (watchdog timer issue)

Record the data that is in menu

item 2.2.2 DIAGNOSTIC

(Configuration mode /

Supervisor menu). Speak to the

supplier.

ERR 101 Current output drift

1

The current output is not

calibrated.

Speak to the supplier to get the

calibration procedure.

1

Hardware error.

Replace the device.

ERR 102 Temperature out of

range

3

The ambient temperature is

outside the given range. This can

cause loss or corruption of data.

Measure the ambient

temperature. De-energize the

device until the ambient

temperature is back in the given

range. If the temperature does

not stay in the correct range,

make sure that there is

insulation around the signal

converter. If this error occurs 2

times, replace the device.

ERR 103 Converter memory

failure

1

The device's hardware is

defective.

Replace the signal converter.

For more data, refer to

How to

turn or remove the signal

converter

on page 43.

ERR 104 Converter voltage

failure

1

The device's hardware is

defective.

Replace the signal converter.

For more data, refer to

How to

turn or remove the signal

converter

on page 43.

Sensor errors

ERR 200 Reference pulse

lost

2

Reference amplitude is less than

reference threshold. This fault

could occur because the device

hardware is defective.

Speak to your supplier to make

sure that the electronics are

still functioning correctly.

Make sure that your installation

has ESD protection. For more

data, refer to

Nozzles on conical

silos

on page 28.

ERR 201 Sensor voltage

failure

1

The device's hardware is

defective.

Do a check of the power supply

at the device terminals. Make

sure that voltage values are in

the specified limits in menu item

2.2.2 DIAGNOSTIC

(Configuration mode /

Supervisor menu). If the voltage

is correct, replace the signal

converter. For more data on

how to replace the signal

converter, refer to

How to turn

or remove the signal converter

on page 43.