Operation, 4 further information on device configuration, 1 quick setup – KROHNE OPTIFLEX 2200 C_F EN User Manual
Page 80
![background image](/manuals/369278/80/background.png)
6
OPERATION
80
OPTIFLEX 2200 C/F
www.krohne.com
07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
6.4 Further information on device configuration
6.4.1 Quick Setup
Use this procedure to change the probe length and give the top and bottom measuring limits.
«
xx » in the illustrations shows that you can change the value or the parameter. Push the
keypad buttons in the correct sequence:
Procedure
Screen
Steps
Description
•
[>>>>], [
] and [>>>>].
Default screen.
Enter configuration mode (2.0.0
SUPERVISOR).
•
[>>>>], [^
^
^
^], [
], [
], [>>>>] and [^
^
^
^].
Enter the password (the default password
is shown). If it is necessary to change the
password, refer to
Function description
on page 74, menu item 2.7.5 PASSWORD.
•
[>>>>]
Push this button to start the quick set-up
procedure.
•
[>>>>] to change the tank height (H).
•
[>>>>] to change the position of the cursor.
•
[
] to decrease the value or [
] to
increase the value.
•
[^
^
^
^] to confirm.
•
[
] or [
] for the selection of the
measurement name (Distance, Level,
Conversion or Ullage Conv.).
•
[^
^
^
^] to confirm.
The manufacturer sets the output function
to “Level” before delivery.
If it is necessary to measure volume,
ullage volume, mass or ullage mass
(Conversion or Ullage Conv.), refer to
How
to configure the device to measure volume
or mass
on page 85.
•
[
] or [
] for the selection of the current
output range (4-20 mA/3.6E, 4-20,
3.8-20.5/3.6E, etc.).
•
[^
^
^
^] to confirm.